Fluke 700LTP-1 Bedienungsanleitung

Fluke pomp 700LTP-1

Lies die bedienungsanleitung für Fluke 700LTP-1 (1 Seiten) kostenlos online; sie gehört zur Kategorie pomp. Dieses Handbuch wurde von 32 Personen als hilfreich bewertet und erhielt im Schnitt 4.6 Sterne aus 16.5 Bewertungen. Hast du eine Frage zu Fluke 700LTP-1 oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage

Seite 1/1
®
PN 2811813 April 2007 © 2007 Fluke Corporation. All rights reserved. Printed in U.S.A. 1
Fluke-700LTP-1
Low-Pressure Test Pump
Instruction Sheet
Introduction
The Fluke-700LTP-1 Low-Pressure Test Pump is a
portable, dual source of vacuum and pressure. Each pump
incorporates a vacuum and pressure selector and fine
adjustment control. The pump has the following
specifications:
Output pressure: 0 to 100 psi / 0 to 6.9 bar
Output vacuum: 0 to -13 psi / 0 to -900 mbar
Exposed Materials: Aluminum, silicon, neoprene,
stainless steel, Buna-N
Adjustment: Fine volumetric pressure and vacuum
adjuster
Dimensions: 150 mm x 100 mm
Weight: 226 grams or .5 lbs (pump only)
Box Contents
Fluke 700LTP-1 Low-pressure Test Pump
1/8 in NPT male/female tee
(2) 1 m hoses
(4) 1/8 in NPT male quick connects
(2) 1/8 NPT female to 1/4 in BSP female
Seal Kit
Instruction Sheet
Contacting Fluke
To contact Fluke or for service, call one of the following
telephone numbers:
USA: 1-888-44-FLUKE (1-888-443-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com.
1
2
8
3
5
6
7
4
eus01f.eps
A
Pressure/vacuum release valve- Turn valve counter-
clockwise for vacuum, clockwise for pressure and
centered for release.
B Fine adjustment control- Turn in or out to increase or
decrease pressure accordingly.
C Master and test port tee
D Quick-fit connectors
E Hose
F 1/8 in NPT female to 1/4 in BSP female fitting
G
Pump Handle. If the pump has not been used for a
while, there may be some resistance on the first
stroke.
H Cleanout ports (2)- See cleanout instructions
Figure 1. Features
Guidelines for Use
1. Connect a pressure calibrator or pressure module to
the quick-fit connector (item D) at the end of one of the
flexible hoses (item E).
2. Connect the unit under test to the quick-fit connector
(item D) at the end of the second flexible hose (item
E F), choosing correct adapters (item ) and seals.
3. Turn the fine adjustment control to mid-range.
4. Set control valve to pressure.
5. Operate handle (item G) until the pressure is close to
the required value.


Produktspezifikationen

Marke: Fluke
Kategorie: pomp
Modell: 700LTP-1

Brauchst du Hilfe?

Wenn Sie Hilfe mit Fluke 700LTP-1 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten