Fluke 700PRV-1 Bedienungsanleitung
Seiten 2 Englisch Pumpe
Lies die Bedienungsanleitung für Fluke 700PRV-1 kostenlos online oder lade das PDF herunter. Diese Anleitung wurde 34 Mal aufgerufen. Frage zu diesem Modell? Stelle eine Frage

PN 2811851
April 2007
©2007 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Introduction
The Fluke-700PRV-1 Pressure Relief Valve is an adjustable
pressure relief valve for the Fluke-700HTP-1 Hydraulic Test
Pump. The valve provides a pressure relief setting range of
50 to 400 bar (725 to 5,800 psi).
The 700HTP-1 pump can develop pressures up to 690 bar
(10,000 psi). Application of too much pressure can damage
some pressure instruments. The pressure relief valves in this
kit provides a method to protect pressure instruments from
accidental overpressure. When the applied pressure reaches
the relief setting of the valve, the valve opens and releases
hydraulic fluid pressure.
Specifications
Preset Value (±
±
±
±±10 %) Relief Setting Range
2900 psi
200 bar
725 to 5800 psi
50 to 400 bar
Pressure Connection: 1/8-in NPT male thread. This mates
with the pressure relief valve port on the 700HTP-1 pump.
Pressure Relief Adjustment Method: Multiturn adjustment
screw to set preload on disc springs.
Repeatability of relief setting:±10 % of nominal setting
Material: 303 stainless steel
Dimensions (each): 1.52 in by 0.62 in diameter (38 mm by
16 mm diameter)
Weight (each): 1.6 oz (46 g)
®
Fluke-700PRV-1
Pressure Relief Valve Kit
Instruction Sheet
Produktspezifikationen
| Marke: | Fluke |
| Kategorie: | Pumpe |
| Modell: | 700PRV-1 |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Fluke 700PRV-1 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten