KERN ORL 94BS Bedienungsanleitung
Lies die bedienungsanleitung für KERN ORL 94BS (1 Seiten) kostenlos online; sie gehört zur Kategorie Messung. Dieses Handbuch wurde von 4 Personen als hilfreich bewertet und erhielt im Schnitt 4.3 Sterne aus 2.5 Bewertungen. Hast du eine Frage zu KERN ORL 94BS oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage
Seite 1/1

Catalogue
1.Introduction
1.Introduction
1
2
2
3
6
7
8
8
9
2.Display and Buttons
2.Display and Buttons
3.Preparations before Operation
3.Preparations before Operating
4.Booting and Measurement
4.Booting and Measurement
5.The Calibration
6.Scales Converting and Temperature Systems
Converting
7.Turn Off
8.Maintenance and Preservation
9.Appendix
Before operating your instrument, please read this
manual properly.
Panel Descriptions
3.1 Install the Battery
3.2 USB Connection
4.1 Booting
4.2 Measurement
5.The Calibration
The meter only supports pure water calibration.
The calibration method is as following:drop 0.4ml
pure water then close the cover to measure.
6.Scales converting and temperature systems converting
6.1Scales Converting
6.2Temperature System Converting
If exceed the temperature limitations,
the signs “HHH” or “LLL” would show.
7.Turn Off
1. After turn on if without any operations for 3 minutes, the
instrument would be automatically turned off.
2. Press “on/ off ” button for 2-3 seconds, the instrument
would be turned off.
8.Maintenance and Preservation
9.Appendix
1. Please clean and wash the sample plate with distilled
water and dry it with soft cleaning cloth or paper towel
after finishing the measuring one kind sample.
2. Never left the remains and residuals of samples
in the sample plate for long time.
3. After finishing measurements of the corrosive
liquid, please clean the sample plate as quick
as possible to avoid the irreparable damage of
the prism and metal surface of the plate.
4. Please use soft cleaning cloth or paper towel to
clean the sample plate to avoid scribing the
prism’s glass.
5. When the dropper and dust-free cloth are not used,
please clean it with clean water and put it in the packing
box after drying.
6. If no using the instrument for a long time,
please remove the battery, and preserved in a
cool and dry environment.
Performance:
The Error Codes Table:
Range
Accuracy Resolution
Brix
R.I.
Temperature
Dimension
Net weight
0.0%~94.0% ±0.1%
0.1%
0.0001
0.1℉
0.1℃
180*100*55mm
365g (excluding battery)
±0.0002
±0.3℃
±0.6℉
1.3330~1.5290
0.0~40.0℃
32.0~104.0℉
Code Instructions
A01
A02
A03
Beyond the scope of calibration
temperature. 1.0 ~40.0 )( ℃ ℃
During calibration, no solution
or non-pure water
This instrument has a hardware failure.
Description of Scales Numbering:
Use the screwdriver to open the battery hatch.
Press button
for 2 seconds,
temperature
unit would be
converted.
This digital refractometer cannot measure any liquid
that is highly corrosive to metal or glass. When
measuring liquids that are corrosive to plastics or react
chemically with plastics, be careful not to drop the
measured liquid onto the shell. Otherwise it will corrode
the shell.
2 3
4
5
6 7
8
9
10
1
Press CAL once
Cyclic Switch
USB lnterface
Connect to USB for charging/connecting to the computer.
Packaging p1-x1 The instructions p1-x1 Dropper p1-x1
Screwdriver p1-x1 USB Line p1-x1 Charger p1-x1
Note :
1. When used outdoors,please avoid strong light
so as not to affect the measurement accuracy.
2. Before dripping into the sample liquid, please
clean the sample plate and Prism with soft clean
cloth or soft paper.
3. Please keep the instrument in a stable status to
measure.
4. Please ensure instrument, environment and
sample are in the same temperature level before
measuring.
After tum on,clear the distilled water and dry the
sample plate, drip 0.3~0.4ml sample then close the
cover to measure.
Cover
USB Port
Charge Lamp/Charging lndicaor
Battery Hatch
LCD Display Screen
Button
Host Display Area
Battery Capaclty
Temperature
Compensation
Refractive Index
Temperature Display Area
Multi-function Display Area
USB Connect
Power on / off
Calibrating “Zero
Point”
Converting Different
Scales&
Temperature Unit
Measurement
Display Areas and Buttons
Press "On/Off"
button to turn on
The multi-function display area
would show the current scale number.
If there is no sample,it will display “dash”
Press the “Read”
one time the meter
will measure one time.
Showing Beyond
Measuring Range
Measuring Result
If press the "Read" button for 2 seconds, the
instrument would make the automatic measurements
upon programmed times (default 15times), the final
value is the average of 15 times’ measurements, After
measurements, it will return to the normal state
and the scales display area would display the average of
the automated measurements
The multifunctional area
displays the number of
remaining automatic
measurements.For each
automatic measurement,
this number is reduced by 1.
Press “CAL” button
for 2-3 seconds till
see the ‘CAL’
flashing .to enter the
calibration status.
LED Flashing Display
Press "CAL" button once again during the ‘CAL’
flashing to start to calibrate. When the calibration is
completed showing as following.
If no any operations for 10seconds the instrument
would return back to booting status.
If fail to complete the calibration, multi-function
display area would show an error code.
Display error code A03,indicating error of sensor
measurement.See the appendix for other error codes .
Press‘Scale’once to
convert the scales
and the values.
Current Scale Number
Switching
Switching
Temperature Unit
The Packing Accessories:
Prism
Stainless Steel
Sample Groove
If is displayed, please charge it.
2024/09 V3.2
We reco end calibrating the refractometer,mm
whe co issi ingn mm on
afte a s ng shockr tro
after longer transp tor
after a change of location with a large temperature
difference
i the device has not b n used fo a long imef ee r t
Always use distilled water and make sure that the
refractometer, he water and the envi nment are a he t ro t t
sa t perature.me em
Produktspezifikationen
Marke: | KERN |
Kategorie: | Messung |
Modell: | ORL 94BS |
Brauchst du Hilfe?
Wenn Sie Hilfe mit KERN ORL 94BS benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Messung KERN
10 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
21 Juni 2025
21 Juni 2025
21 Juni 2025
20 Juni 2025
Bedienungsanleitung Messung
- SKS
- Softing
- Greisinger
- Multimetrix
- Phoenix Contact
- Emko
- P3 International
- Megasat
- Adwa
- Steinberg
- ORNO
- UHoo
- Festo
- Duro Pro
- Alecto
Neueste Bedienungsanleitung für -Kategorien-
10 Juli 2025
10 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025
9 Juli 2025