Manometer
USER MANUAL
Description of the device
1. OK key
3. Select key
4. On/Off/Esc key
5. Hold/min/max key
MP 120
Technical specications
General features
*All the accuracies indicated in this technical datasheet were stated in laboratory conditions, and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation.
Measuring elementPiezoresistive sensor
Tolerated overpressure250 mbar
ConnectorØ 6.2 mm threaded connectors made of nickelled brass
Display
4 lines, LCD technology. Dimensions 50 x 36 mm.
2 lines of 5 digits with 7 segments (value)
2 lines of 5 digits with 16 segments (unit)
HousingABS, protection IP54
Keypad5 keys
European directives2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS II; 2012/19/EU WEEE
Power supply4 batteries AAA LR03 1.5 V
Battery life180 hours
AmbienceNeutral gas
Conditions of use (°C,%RH, m)From 0 to +50 °C. In non condensing conditions. From 0 to 2000 m.
Storage temperatureFrom -20 to +80 °C
Auto shut-offAdjustable from 0 to 120 minutes
Weight220 g
ParametersMeasuring unitsAccuracyMeasuring rangeResolution*
PressurePa, mmH
2
O, inWg, daPa, m/s, fpm±0.5% of reading ±2 PaFrom -1000 to +1000 Pa1 Pa
Air velocitym/s, fpm, km/h
From 2 to 5 m/s: ±0.7 m/s
From 5 to 40 m/s: ±0.5% of reading
±0.3 m/s
From 0 to 40 m/s0.1 m/s