Lies die bedienungsanleitung für Extech 407910 (9 Seiten) kostenlos online; sie gehört zur Kategorie Messgeräte. Dieses Handbuch wurde von 28 Personen als hilfreich bewertet und erhielt im Schnitt 4.7 Sterne aus 14.5 Bewertungen. Hast du eine Frage zu Extech 407910 oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage
Seite 1/9

User's Guide
Heavy Duty
Differential Pressure Manometer
Model 407910

407910-EN v2.3 3/14 2
Introduction
Congratulations on your purchase of the Extech Manometer Model 407910. This device has a
differential input with convenient quick-disconnect fittings. Eight units of measure are selectable by
front panel pushbutton. Advanced features include Data hold, MAX/MIN record/recall, and RS-232
PC interface. Properly used, this device will provide years of reliable service.
Specifications
General Specifications
Circuit Custom one-chip LSI microprocessor circuit
Display 0.6" (15mm) 6000 count LCD with function indicators
Units of Measure mBAR, psi, kg/cm
2, mm Hg, inches Hg, meter H2O, inches H2O and
Atmospheres
Input circuit Differential inputs (P1 and P2)
Data Hold Freezes displayed reading
Sensor Built-in piezoelectric sensors
Zero Adjust Front panel push-button
Sample rate 0.8 sec. (approx.)
Compatibility Air or non-corrosive and non-ionized gas
Auto Power OFF Meter turns off after approx. 10 minutes to conserve energy
Data Output RS-232 PC serial interface
Ambient conditions 32 to 122
oF (0 to 50oC); < 80% RH
Battery power 9V alkaline battery
Power Consumption 8mA DC (approx.)
Dimensions 7.1 x 2.8 x1.3" (180 x 72 x 32 mm)
Weight 0.46 lbs (210 g) meter only
Accessories Quick-disconnect fittings (2), rubber jacket, and 9V battery
Range Specifications
Range/Units (Max.) Resolution Accuracy
± 2000 mbar 1
± 2% full scale @ 23±5
oC
Note: Accuracy includes linearity,
hysteresis and repeatability effects
± 29 psi 0.01
± 2.040 Kg/cm2 0.001
± 1500 mm Hg 1
± 59.06 inches Hg 0.02
± 20.40 meters H2O 0.01
± 802 inches H2O 0.5
± 1.974 atmospheres 0.001

407910-EN v2.3 3/14 3
Meter Description
1. LCD Display – Indicates the measurement data, unit of measure, and user alert symbols as
described in this manual
2. Push-buttons:
POWER – Press to turn the meter ON or OFF
HOLD – Press to freeze the displayed reading
ZERO – Press to null the display (with fittings disconnected)
RECORD/RECALL – Press to access the MIN/MAX recording mode as described later in
this manual
UNIT – Press to select the unit of measure
3. 0.16” (4mm) Quick-disconnect fittings (supplied)
4. P1 Input – 0.16” (4mm) accepts quick disconnect connector
5. RS-232 PC interface jack (3.5mm) – For use with data acquisition software/hardware interface
kit model 407001
6. P2 input – 0.16” (4mm) accepts quick disconnect connector
7. Rubber protective holster (removable)
8. Battery compartment (on rear) – Remove rubber jacket to access compartment
4
5
6
1
2
3
8
7
Produktspezifikationen
Marke: | Extech |
Kategorie: | Messgeräte |
Modell: | 407910 |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Extech 407910 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Messgeräte Extech
12 September 2024
6 September 2024
6 September 2024
6 September 2024
6 September 2024
6 September 2024
6 September 2024
6 September 2024
6 September 2024
3 September 2024
Bedienungsanleitung Messgeräte
- Toolland
- Skil
- TrueLife
- Vonroc
- Grohe
- Qubino
- SensoProtect
- Seaward
- Uni-Trend
- Eastron
- Wago
- ZKETECH
- Vemer
- TFA Dostmann
- Adwa
Neueste Bedienungsanleitung für -Kategorien-
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024