Lies die bedienungsanleitung für Kyoritsu KEW 8031F (1 Seiten) kostenlos online; sie gehört zur Kategorie Messung. Dieses Handbuch wurde von 12 Personen als hilfreich bewertet und erhielt im Schnitt 3.6 Sterne aus 6.5 Bewertungen. Hast du eine Frage zu Kyoritsu KEW 8031F oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage

Seite 1/1
The following symbols are used and marked on the
instrument and in the instruction manual. Please carefully
check before starting to use the instrument.
Must refer to the Instruction Manual to protect humans
and devices.
Instrument with double or reinforced insulation.
AC
symbol indicates separate collection for electrical and
electronic equipment.
DANGER
Never make measurement on a circuit in which the
earth potential exceeds 600V to avoid electrical
shocks.
Do not make measurement when thunder is rumbling.
If the instrument is in use, stop the measurement
immediately and remove the instrument from the
measured object.
Put insulated protective gears when there is a danger
of electrical shock hazard.
Do not attempt to make measurement in the presence
of flammable gasses, fumes, vapor or dust. Otherwise,
the use of the instrument may cause sparking, which
can lead to an explosion.
Never attempt to use the instrument if its surface or
your hand is wet.
The instrument should be used only in its intended
applications or conditions. Otherwise, safety functions
equipped with the instrument do not work, and
instrument damage or serious personal injury may be
caused.
WARNING
Never attempt to make any measurement, if the
instrument has any structural abnormality such as
cracked case and exposed metal part.
First, connect the Tester to Three-phase system and
then press the push switch button.
If the cables are accidentally disconnected, release
your hand from the push switch button and stop
measurement.
PHASE INDICATOR
KEW 8031 / KEW 8031F
1. Safety Warnings
This instrument has been designed and tested according
to IEC Publication 61010: Safety Requirements for
Electronic Measuring Apparatus. This instruction manual
contains warnings and safety rules which must be
observed by the user to ensure safe operation of the
instrument and retain it in safe condition. Therefore, read
through these operating instructions before starting using
the instrument.
WARNING
Read through and understand instructions contained
in this manual before starting using the instrument.
Save and keep the manual handy to enable quick
reference whenever necessary.
The instrument is to be used only in its intended
applications.
Understand and follow all the safety instructions
contained in the manual. Failure to follow the
instructions may cause injury, instrument damage
and/or damage to equipment under test. Kyoritsu is
by no means liable for any damage resulting from the
instrument in contradiction to this cautionary note.
The symbol
indicated on the instrument means that
the user must refer to related parts in the manual for safe
operation of the instrument. Be sure to carefully read the
instructions following each symbol in this manual.
DANGER
is reserved for conditions and actions that
are likely tocause serious or fatal injury.
WARNING
is reserved for conditions and actions
that can cause serious or fatal injury.
CAUTION
is reserved for conditions and actions
that can cause minor injury or instrument
damage.
WARNING
Don't touch the disconnected cables while the push
switch button is being pressed down.
Stop using the test lead if the outer jacket is damaged
and the inner metal or color jacket is exposed.
Do not install substitute parts or make any modification
to the instrument. Return the instrument to Kyoritsu or
your distributor for repair or re-calibration.
CAUTION
Even if all missing phase indicators are off, one phase
might be live: care should be taken to avoid getting
electric shock.
Max continuous measurement time differs depending
on voltages. Please refer to continuous measurement
time written in this document and measure each
voltage within each limited time.
Do not expose the instrument to the direct sun,
extreme temperatures or dew fall.
Use a damp cloth and detergent for cleaning the
instrument. Do not use abrasives or solvents.
This instrument isn't dust & water proofed. Keep away
from dust and water.
Choose and use the test leads and caps that are
suitable for the measurement category. When the
instrument and the test lead are combined and used
together, whichever lower category either of them
belongs to will be applied.
Keep your fingers and hands behind the protective
fingerguard during measurement.
Measurement categories(Over-voltage categories)
To ensure safe operation of measuring instruments, IEC
61010 establishes safety standards for various electrical
environments, categorized as o to CAT IV, and called
measurement categories. Higher-numbered categories
correspond to electrical environments with greater
momentary energy. So a measuring instrument designed
for CAT III enviroments can endure greater momentary
energy than one desined for CAT II.
3. Instrument Layout4. Specifications
Nominal system voltageUn
AC230V
Voltage ranges
AC110V to 600V
Rated frequency
50Hz / 60Hz
Continuous operation time
Continuous up to AC280V
40 minutes or less from AC281V to AC300V
20 minutes or less from AC301V to AC400V
10 minutes or less from AC401V to AC500V
5 minutes or less from AC501V to AC600V
Safety Standards
IEC 61010-1, IEC 61010-2-030
Measurement category CAT IV 300V,CAT III 600V
Pollution degree 2
IEC 61010-031
IEC 61557-1, IEC61557-7
Standards : IEC 61326-1 (EMC), IEC 60529 (IP30)
Environmental standards : IEC 50581 (EU RoHS)
Location for use
Altitude 2000m or less, Indoor use
Withstand Voltage
AC6300V(rms) for 5seconds
Dimensions
106(L) x 75 (W) x 40 (D) mm (Body)
1.3m (Test lead)
Weight
Approx. 350g
Fuse : 0.5A / 600V (F) 6.3X32mm (only KEW8031F)φ
Accessories
Instruction manual
Carrying Case
5. Operating instructions
WARNING
First, connect the Tester to Three-phase system and
then press the push switch button.
If the cables are accidentally disconnected, release
your hand from the push switch button and stop
measurement.
Don't touch the disconnected cables while the push
switch button is being pressed down.
CAUTION
Even if all missing phase indicators are off, one phase
might be live: care should be taken to avoid getting
electric shock.
Max continuous measurement time differs depending
on voltages. Please refer to continuous measurement
time written in this document and measure each
voltage within each limited time.
DISTRIBUTOR
(1) Connect colour coded alligator clips or prods to the
terminals of a 3-phase power source where a rotating
electrical machine such as a motor will be connected or
input to a building.
(2) Press the push switch button located on top of the
unit. Keep this button pressed during phase sequence
or open phase check. When the push switch button is
released it immediately goes off.
(3) Make sure that all of the three lamps for phase check
are on. If so, there is no open phase. When any of the
three lamps is Not on there is open phase.
Open phase check Open phase on terminal
Lamp "L1" is not on where Red alligator clip
is connected.
Open phase check Open phase on terminal
Lamp "L2" is not on where Whight alligator clip
is connected.
Open phase check Open phase on terminal
Lamp "L3" is not on where Blue alligator clip
is connected.
When the open phase check lamps are not on the
rotating disc does not turn.
(4) Check the rotating direction of the inside disc through
the phase sequence indication window.
When the rotating disc turns counter-clockwise alternate
the connection of the two of the three alligator clips.
Then, the rotating disc will turn clockwise.
When the rotating disc turns clockwise phase sequence
isL1,L2 and L3 in order of the power source terminals
where the Red, Whight and Blue alligator clips are
connected.
1-22 92-2429
Kyoritsu reserves the rights to change specifications
or designs described in this manual without notice and
without obligations.
O(None,Other):Circuitswhicharenotdirectly
connectedtothemainspower
supply.
CATII:Primaryelectricalcircuitsofequipment
connectedtoanACelectricaloutletby
apowercord.
CATIII:Primaryelectricalcircuitsofthe
equipmentconnecteddirectlytothe
distributionpanel,andfeedersfromthe
distributionpaneltooutlets.
CATIV:Thecircuitfromtheservicedrop
totheserviceentrance,andtothe
powermeterandprimaryovercurrent
protectiondevice(distributionpanel).
2. Features
●TwoFunctionsinOneUnit
 KEW8031andKEW8031Fisdesignedto
checkphasesequence.Lampsprovidedon
theunitwillalsotellyouifaphaseisopen.
●HighlyReliable
 
Cancheckawiderangeof3-phasepower
sourcefrom110Vto600V.Sealedagainst
dust,theunitensurestrouble-freeperformance.
●FunctionalDesign
 Small,Lightweightandportable.Designedfor
maximumeaseofoperationandruggedness.
●SafetyDesign
 Noexposedmetalparts.Safetyfeaturesare
incorporatedincludingtheinstantpushbutton
switchoperation.
INSTNMALRUCTIOANU
KEW8031
KEW 8031F
AgatorClipsetlli
RED
WHITE
BLUE
AgatorClipsetlli
RED
WHITE
BLUE
KEW8031
KEW 8031F
Wiring diagram
Protectefigerguard:iv
Whentheinstrumentandthetestleadarecombinedand
usedtogether,whheverlowercategoryeitherofthemic
belongs to wlbe apped.illi
How to replace the fuse(W8031F)KE
Turn the probe tospt itli
Fuse 0.5A/600V(F)
φ6.3X32mm
It is a part providing protection
againstelectrical shock and
ensuring the minimumrequired
air and creepage distances.
O: Device which is
not directly
connected to the
mains power supply


Produktspezifikationen