Lies die bedienungsanleitung für YSI Adapter Multilab BNC (10 Seiten) kostenlos online; sie gehört zur Kategorie Nicht kategorisiert. Dieses Handbuch wurde von 4 Personen als hilfreich bewertet und erhielt im Schnitt 3.6 Sterne aus 2.5 Bewertungen. Hast du eine Frage zu YSI Adapter Multilab BNC oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage

Seite 1/10
OPERATIONS MANUAL
ba76144e01 02/2013
4010-2/3 pH Adapter
DIN / BNC
PH ADAPTER FOR MULTILAB 4010-2 AND MULTILAB 4010-3
2 ba76144e01 02/2013
4010-2/3 pH Adapter
For the most recent version of the manual, please visit www.ysi.com.
1725 Brannum Lane
Yellow Springs, OH 45387 USA
Tel: +1 937-767-7241
800-765-4974
Email: environmental@ysi.com
Internet: www.ysi.com
Copyright © 2013 Xylem Inc.
4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC General information
3
ba76144e01 02/2013
General information
The 4010-2/3 pH Adapter enables the connection of an analog pH electrode
(with or without temperature sensor) to the digital meters MultiLab 4010-2
and 4010-3.
Two models of the adapter are available:
Automatic adapter
recognition
The measuring electronics with the stored adapter data is in the 4010-2/3 pH
Adapter. The data include, among other things, the adapter type and series
number. With each calibration, the calibration data is written in the adapter
and the calibration history is recorded. The data are called up by the meter
and used for measurement and documentation of measured values.
If the adapter firmware is enhanced by YSI, it can be updated via the meter.
Note
More detailed information on the automatic adapter recognition or sensor rec-
ognition is given in the operating manual of the meter.
Calibration data
and calibration
history
The meter uses for measurement the calibration data stored in the adapter,
regardless of the mounted electrode.
Note
Therefore, you always have to recalibrate after exchanging the pH electrode.
Otherwise, the IDS pH sensor will work with the calibration data of the
previously calibrated electrode. This could lead to incorrect measured
values. When evaluating the calibration history, please note that in between
calibrations, the electrode may have been exchanged.
Connection of the analog electrode Adapter
DIN 4010-2/3 pH Adapter DIN
BNC 4010-2/3 pH Adapter BNC