Lies die bedienungsanleitung für YSI TruLine Ammonia Electrode (2 Seiten) kostenlos online; sie gehört zur Kategorie Nicht kategorisiert. Dieses Handbuch wurde von 26 Personen als hilfreich bewertet und erhielt im Schnitt 4.3 Sterne aus 13.5 Bewertungen. Hast du eine Frage zu YSI TruLine Ammonia Electrode oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage

Seite 1/2
Required Equipment
Ionmeter(preferred) orpH/mVmeter.
Washbottlewithdistilledordeionizedwater.
. Severalcleanbeakers(glass, 150 mL preferred)
. Pipette for measuring 10 mL or less
Graduated cylinder (100 mL)
* FillSolutionYSI #400367
() 1000mg/L ppmAmmoniaStandardYSI #400370*
) 100mg/L (ppmAmmoniaStandard#400369 YSI
Replacement MembraneModules(pk/4)YSI#400366**
*Small bottles included with ammonia kit (YSI #400365)ISE
**Two Membrane modules included with ammonia ISE kit.
Overview
The400365 YSI a - AmmoniaIonSelectiveElectrodeisgas
sensingelectrodeformeasuringammonia (NH
3
) inaqueous
samples.beusedthatbufferssample Itmust withanISA the
>pH11 tocause theammoniainthesampletobecome
gaseous.AmmoniaElectrodebehindan The a ISEispH
ammoniagas permeable membranetheammoniagas
causespHshiftthatcorrelated a intheFillSolution is to
ammoniaconcentration based on the resultof calibrations .
Technology: Gassensingtechnology
Size and material:
MembraneBody (ABS)12mmOD x 158 L mm
Cap x (ABS)16mmOD37 L mm
Cable(Coax00cm 1type); BNC Connector
a -Reference: ThepHelectrodeinsidehassingle
junction,Ag/AgCl, annularceramic junction.
- Features: Replaceable, preformedmembrane module.
Thea pH ISAhas indicator that indicates the pHisin
rangebya (>11) thepresenceof bluecoloration.
Range: 0.02- 17,000mg/L
Reproducibility: ±2%
@ Slope: 54to 59mv/decade25°Ctypical
pH range: >11
0 Operating Temperature:to 50°C
Interferences: Volatileamines
Applications/Notes: Popular for wastewater,
aquaculture, agriculture, and industrial applications.
Electrode Preparation
1. The glass pH sensor and reference junction of the pH
electrode should be kept conditioned during transit and
storage. theofthesolutionUnscrew cap storage bottle
(Soakerandelectrode. Bottle) removethepH
2. Fill a MembraneModulewithFillSolution
a) Add - s 1520dropof Fill SolutiontotheMembrane Module
b) TapModuleensurebubbles theMembrane to no are
present . inthefill solution
c) ThreadMembraneModulethe the slowly into Electrode
Cap. If theModulewasoverfilled, theexcessFill Solution
will leak out. AvoidrupturingtheMembraneModulewith
toomuch Fill Solution.
3. Preferably, allow thepH electrodetoconditionwithinthe
filled 2 MembraneModuleforhourspriortouse.The
Membrane Module can be conditioned by placing in a low
to mid-range standard for at least 15 minutes prior to use.
Calibration
1. Connect the assembled ammonia electrode to a meter with
BNC input. Meters with an ISE mode can directly display ion
concentration (e.g. display in mg/L). Ensure the meter is
setup for ammonia and be sure to follow the calibration
procedure in the meter user manual.
Alternatively, a meter with a mV mode can be used, although
the instrument will not display the result of calibration (i.e.
calibration curve will need to be created by the user).
2. For highest accuracy, use a meter with automatic
temperature compensation (ATC) and connect a temperature
sensor.
3. Prepare Standards:
a)
At least 2 standards that bracket the expected sample
range must be used. As an example, prepare 1 mg/L and
10 mg/L standards if the expected sample is ~5 mg/L.
Some meters allow up to a 7 point calibration.
b) There mustbe at least one order of magnitude
difference between high and low standards (e.g. 10
mg/L and 100 mg/L).
c) If standards span more than one order of magnitude
(e.g. 1 mg/L and 100 mg/L), it is best to prepare at least
one mid-range standard (e.g. 10 mg/L).
d) Each standard should have a volume of 100 mL and
should be placed in a 150 mL glass beaker.
e) Standards should be fresh and carefully prepared. It is
best to use a pipette when measuring small volumes
(10 mL or less) of a more concentrated standard.
f) Use the equation C1*V1=C2*V2 when preparing diluted
standards from a more concentrated standard solution.
As an example, if using the YSI 100 mg/L standard as a
stock solution (YSI #400369) to make 100 mL of a 10
mg/L standard: C1=100 mg/L, V1=?, C2=10 mg/L, and
V2=100 mL. V1=(C2*V2)/C1 =10 mL.
Therefore, 10 mL of the 100 mg/L stock standard should
be added to 90 mL of DI water (i.e. solution should be 100
mL) to make a 10 mg/L standard.
Note:It is not recommended to use a very concentrated
standard (e.g. 1000 mg/L) to make a very dilute standard
(e.g. 0.1 mg/L).
4. Place the electrode in the lowest concentration standard and
stir at a constant rate using a stir bar and stir plate, although
the stirring speed should be limited to minimize loss of
ammonia gas. Use the same stirring rate throughout
calibration.
If you do not have a stir plate, it is best to swirl the solution
or use a stir rod once ISA is added.
5. Add 2 mL of ammonia ISA to the first standard:
a) Using a pipette is recommended. The standard solution
should have a pale blue color that persists after ISA is
added.
b) Allow the solution to stir for 1 minute and begin
calibration. Calibration is time sensitive, as ammonia
diffuses from the sample~4 minutes after ISA is added.
Therefore, ensure calibration with the standard is
complete 1 to 4 minutes after ISA is added. This timing
is critical!If calibration using the standard is
unsuccessful within this time limit, a fresh standard will
n eed to be prepared.
6. After calibrating with the first point, finish calibrating using
steps #4 and #5 for the remaining calibration points.
C alibrate in order of increasing concentration.
7. Evaluate electrode slope:
a) View the calibration record on the ISE meter, or create
a curve if using a meter with only a mV mode.
b) For highest accuracy, the ammonia electrode slope
should be between 52 mV/decade and --60 mV/decade.
8. Recalibrate often:
a) The ammonia electrode should be calibrated at the
beginning of each day.
b) For best accuracy, it is best to calibrate the ammonia
electrode every 10 samples. If this calibration interval is
too frequent, at least verify your calibration result every
2 hours by preparing a fresh low to midrange standard, -
adding ISA, and verifying thereading. If the mV reading
has changed ~3 mV compared to the reading in that
standard during calibration, you will need to recalibrate
the electrode.
Measurement
Samples must be prepared like standards. Follow steps #3-5
of the calibration procedure when preparing samples and
taking a measurement.
In summary, 100 mL of sample should be collected and 2 mL
of ISA added. The sample should be fresh and the same
stirring rate using during calibration should be utilized. A
measurement can be taken 1 to 4 minutes after ISA is added.
The temperature of measurement solutions should be as
close as possible to solution temperatures during calibration.