Lies die bedienungsanleitung für YSI Multilab IDS 4210W ORP (12 Seiten) kostenlos online; sie gehört zur Kategorie Nicht kategorisiert. Dieses Handbuch wurde von 2 Personen als hilfreich bewertet und erhielt im Schnitt 3.5 Sterne aus 1.5 Bewertungen. Hast du eine Frage zu YSI Multilab IDS 4210W ORP oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage

Seite 1/12
OPERATIONS MANUAL
ba76138e05 06/2018
4210W
4211
ORP ELECTRODE
2 ba76138e05 06/2018
4210W / 4211
Contact YSI
Yellow Springs, OH 45387 USA
Tel: +1 937-767-7241
800-765-4974
Internet: www.ysi.com
Copyright © 2018 Xylem Inc.
For the most recent version of the manual, please visit
www.ysi.com
.
4210W / 4211 General information
3
ba76138d05 06/2018
1 General information
Automatic sensor
recognition
The sensor electronics with the stored sensor data are in the connecting
head of the electrode. The data include, among other things, the sensor type
and series number. The data are recalled by the meter when the sensor is
connected and are used for measurement and for measured value documen-
tation.
The digital transmission technique guarantees the failure-free communica-
tion with the meter even with long connection cables. The sensor firmware
can be updated via the meter.
2 Commissioning, measuring, checking
2.1 Commissioning
Scope of delivery Electrode 4210W or 4211
Operating manual
Commissioning Prepare the electrode for measuring as follows:
Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop-
per or a slider.
The refilling opening must always be open during measurement!
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Connect the sensor to the meter.
The sensor is immediately ready to measure.
Please keep the watering cap. It is required for the electrode to
be stored. Always keep the watering cap clean.
4211 via the sensor cable to a free IDS connec-
tor on the meter
Refilling
opening