Lies die bedienungsanleitung für Wachendorff TA327401 (2 Seiten) kostenlos online; sie gehört zur Kategorie Messung. Dieses Handbuch wurde von 9 Personen als hilfreich bewertet und erhielt im Schnitt 3.6 Sterne aus 5 Bewertungen. Hast du eine Frage zu Wachendorff TA327401 oder möchtest du andere Nutzer dieses Produkts befragen? Stelle eine Frage

Seite 1/2
INTRODUCTION
Thanks for choosing a Wachendorff Prozesstechnik device.
The tachometer TA327401 allows to read the frequency (max 100 kHz) of a
signal from single or double (bidirectional encoder) input.
2 universal digital inputs are available (NPN/PNP ) for
external commands like output activation or Hold/ Stop current visualization;
one input is also analogue in order to allow setpoint modification by external
potentiometers.
/potential free contact
WIRING DIAGRAMWIRING DIAGRAM
Potentiometer:
1
2
3
4
To modify Set1 or Set2 by external potentiometer follow the steps below:
- use potentiometers 0 kOhm to 5/10 kOhm
- connect cursor to pin I3; a wrong connection may damage the potentiometer and
lead to lock of the device.
- accuracy on input is max 1000 points, therefore set the parameters "Upper limit"
and "Lower limit" with a max difference of 1000 units.
(Ex.: LoS1 to 50,0 and uPS1 to 150,0 to modify setpoint value related to Set1
between 50 and 150 steps with steps of one tenth). Greater differences would make
unstable the less significant digit.
- To calibrate the scale of potentiometer enter the configuration mode and select:
Hin.3 as Pot Fin.3 as Set1 or Set2 P.tAr as Enable
Exit configuration mode and place potentiometer at minimum level and press key,
then place potentiometer at max level and press key: the device automatically exit
the calibration procedure.
N.B.:Aswitch-off of the device would interrupt the calibration.
PRESS
for 3 seconds
Display 1 shows
with 1st digit blinking, while
Display 2 shows
DISPLAY DO
or
Modify blinking digit, pass to the
next one pressing
Enter password
to confirm
Display shows first parameter of
configuration table
1
2
3
Scroll parameters
4
or
Increase or decrease value on display
pressing and an arrow key
Enter the new data
that will be stored
when releasing the keys
5
+ or
End of configuration,
the device exits from programming mode.
6
CONFIGURATION PARAMETER MODIFICATIONCONFIGURATION PARAMETER MODIFICATION
Visualizes SETPOINT 1 / 2
or
Modifies selected SET
Selects chosen digit
1
2
2a
Modifies blinking digit
of selected SET
3a
or
PRESS DISPLAY
SETPOINT MODIFICATIONSETPOINT MODIFICATION
SIZE AND INSTALLATIONSIZE AND INSTALLATION
U1
MEMORY C.121MEMORY C.121
123
Frontal panel cut-out
28.5 mm x 70.5 mm
2mm
Suggested thickness
÷8
77 mm
35 mm
53 mm
7
42 mm pins
Memory Card
CN1
1 6
Memory Card (optional)
with battery SFUR0KIT
Gasket for 32 x 74
Parameters and setpoint values can be copied from one device to another using
the Memory card
There are two methods:
insert the memory card .
On activation display 1 shows and display 2 shows
The device loads the new data and starts again
The memory card is equipped with an internal battery with a life of about 1000 uses.
Insert the memory card and press the programming button.
When writing the parameters, the LED turns red and on completing the procedure
it changes to green. It is possible to repeat the procedure.
To the memory card values, follow the procedure described in the first
method, setting display 2 to so as not to load the parameters on controller.
Enter configuration and .
Exit configuration. Changes are saved automatically.
With the device connected to the power supply
when the controller is off
With the controller disconnected from the power supply:
change at least one parameter
MemO
update
.
(Only if the values stored on Memory Card are correct).
By pressing the key display 2 shows
Confirm using the key .
.
Attention: Perform first an update of the programm module.
>
>
UPDATING MEMORY CARD.
MEMORY CARD (optional)MEMORY CARD (optional)
PRESS
for 3 seconds
Display 1 shows
with 1st digit blinking, while
Display 2 shows
DISPLAY DO
or
Modify blinking digit, pass to the
next digit pressing
Enter password
to confirm
The device loads
default settings
Switch the device off
and restart it
1
2
3
LOADING DEFAULT SETTINGSLOADING DEFAULT SETTINGS
PARAMETERS LISTPARAMETERS LIST
Disabled
Default
Enabled
P-10 Potentiom. Tarature Potentiometer calibration procedure
Disable
Enable
P-01 Clock Input
CLOCK INPUT CONFIGURATION
Input signal selection
Input signal on I1
Default
I1
Encoder
Input signal on I1 and I2 (bidirectional encoder)
P-02 Hardware input 1
INPUT CONFIGURATION
Input 1 hardware configuration
NPN (not available on input 3)
Default
PNP
TTL
Potentiometer (available only for input 3)
P-03 Hardware input 2 Input 2 hardware configuration
P-04 Hardware input 3 Input 3 hardware configuration
P-05 Filter Input 1 Input 1 hardware filter configuration
Input hardware filter desabled
Default
Input hardware filter enabled (22nF)
Off
On
P-06 Active State Input 2 Input 2 active state
P-07 Active State Input 3 Input 3 active state
High level
Default
High Level
Low Level
Low level
Disabled
Default
Enable / Desable tachometer outputs
Hold visualized tachometer value
P-09 Function Input 3 Function associated to Input 3
Set1 setting by potentiometer
Set2 setting by potentiometer
P-08 Function Input 2 Function associated to Input 2
Disabled
Default
Disable
Display max peak
Max. registered peak visualization (reset by UP+DOWN key)
P-11 Function Key UP Function associated to key UP (up arrow)
Disabled
Default
Disable
Display min peak
Min. registered peak visualization (reset by UP+DOWN key)
P-12 Function Key DOWN Function associated to key DOWN (down arrow)
P-13 Power-off Memory
BACKUP MEMORY CONFIGURATION
Power-off memory
No peak value stored at switch-off
Default
Disable
Min Peak
All Peak
Max Peak
Max. and Min. peak values stored at switch-off
Maximum peak value stored at switch-off
Minimum peak value stored at switch-off
If enabled maximum peak function, maximum peak
value obtained is visualized.
If enabled minimum peak function, minimum peak
value obtained is visualized.
If enabled peak function, minimum and maximum peak
value will initialize to current tachometer value.
1
2
3
PRESS
DISPLAY
MAXIMUM AND MINIMUM PEAK FUNCTIONMAXIMUM AND MINIMUM PEAK FUNCTION
and
Disable
Out Enable/Disable
Hold (only for I3)
Set1 (only for I3)
Set2 (only for I3)
NPN
PNP
TTL
Potent.
No decimal digit visualization
Default
0
0.0
0.00
0.000
1 decimal digit visualization
2 decimal digits visualization
3 decimal digits visualization
Default
P-18 Decimal Point Tachometer value visualization format
OUTPUT ENABLE CONFIGURATION
P-29 Output Enable Outputs enabled
Tachometer outputs always enabled
Default
Always enable
Automati enable
Enable by input
Outputs enabled automatically
Tachometer outputs enabled by digital inputs
SETPOINT CONFIGURATION
Setpoint value not visualized
Default Set2
Setpoint value visualized
P-23 Display Set 1 Setpoint 1 display selection
P-26 Display Set 2 Setpoint 2 display selection
Disable
Visualized
Modifiable
Setpoint value visualized and modifiable
Default Set1
P-24 Lower Limit Set 1 Set 1 minimum value (0...9999)
P-27 Lower Limit Set 2 Set 2 minimum value (0...9999)
Default 0
Default 0
P-25 Upper Limit Set 1 Set 1 maximum value (0...9999)
P-28 Upper Limit Set 2 Set 2 maximum value(0...9999)
Default 999
Default 999
MEASURE UNIT CONFIGURATION
Set each of 4 digits as chosen
Default ----
P-19 Measure Unit 1 Setting digit 1 of displayed measuring unit
P-20 Measure Unit 2 Setting digit 2 of displayed measuring unit
Edit digits
P-21 Measure Unit 3 Setting digit 3 of displayed measuring unit
P-22 Measure Unit 4 Setting digit 4 of displayed measuring unit
Default
Active output with low deviation
Active output inside band
Active output out of band
P-30 Logic Output Mode1
P-34 Logic Output Mode2
Active output with high deviation
Tachometer logic output mode 1
Tachometer logic output mode 2
TACHOMETER LOGIC OUTPUT MODE CONFIGURATION
Default
Automatic
Latch output (clear by FNC key)
Pulse 0.1 sec
to
Pulse 99.9 sec
Latch output, reset by FNC
0,1 sec output impulse duration
P-33 Output 1 Duration
P-37 Output 2 Duration
Automatic output duration
Tachometer logic output 1 duration
Tachometer logic output 2 duration
99,9 sec output impulse duration
Default
0.0 sec
to
999.9 sec
Defines logic output activation delay.
P-31 Activation Delay 1
P-35 Activation Delay 2
Logic output 1 activation delay
Logic output 2 activation delay
to 999.9 sec.
Setting range from 0.0 sec
Default
0.0 sec
to
999.9 sec
Defines logic output deactivation delay.
P-32 Deactivation Delay 1
P-36 Deactivation Delay 2
Logic output 1 deactivation delay
Logic output 2 deactivation delay
to 999.9 sec.
Setting range from 0.0 sec
P-38 Output Q1 Setup Relay Q1 output setting
P-39 Output Q2 Setup Relay Q2 output setting
Disabled output
Default 1
Disable
Logic Out 1 n.o.
Logic Out 1 n.c.
Logic Out 2 n.o.
Logic Out 2 n.c.
Logic output 1 on n.o. contact
Logic output 1 on n.c. contact
Logic output 2 on n.o. contact
Logic output 2 on n.c. contact
Default 2
OUTPUT CONFIGURATION
High Deviation
Low Deviation
Inside Band
Out of Band
P-17 Pulse in Unit Impulses on visualized unit
Number of impulses for single unit. For example,
Default
1 pulse
...
9999 pulse
in speed measurement, it indicates how many
Impulses correspond to a full revolution.
P-14 Minimum Input Frequency
CLOCK INPUT CONFIGURATION
Lower frequency visualized
For lower frequency values 0 is visualized on display.
Default
0.1 Hz
...
10.0Hz
This parameter forces max. refresh time of display
from 100 to 0.1 sec.
P-15 Software Filter Sampling frequency software filter
No software filter on reading
Default
off
0.01 sec
...
1.00 sec
Mean realized on samplings done within time set
in this parameter. Display will be updated according to
this time range.
DISPLAY CONFIGURATION
P-16 Timebase Visualization time base
Visualized value referred to the second
Default
sec
min
hour
Visualized value referred to the minute
Visualized value referred to the hour
0.01 Hz
...
0.09Hz
99.99 pulse
...
0.01 pulse
Read carefully the safety guidelines and programming instructions
contained in this manual before using/connecting the device.
Disconnect power supply before proceeding to hardware settings or
electrical wirings.
Only qualified personnel should be allowed to use the device and/or
service it and in accordance to technical data and environmental
conditions listed in this manual.
Do not dispose electric tools together with household waste materials in
observance of European Directive 2002/96/CE
POTENTIOMETER 0 to 5/10 Kohm
Resolution 1000 steps.i
GND
POSITIVO
Fase Z
Encoder
Fase B
Fase A
PNP
NPN
PNP
NPN
Proximity
Sensor PNP
3 wires
Proximity
Sensor PNP
2 wires
+-
OUT
OUT
5A 230V
Resistive
1/2HP
Q1
5A 230V
Resistive
1/2HP
Q1
5A 230V
Resistive
1/2HP
Q2
5A 230V
Resistive
1/2HP
Q2
TECHNICAL DATATECHNICAL DATA
Operating
Conditions
Sealing
Material
Digital
Inputs
Outputs
OUT 24V
Back-UP
Power Supply
Operating temperature: 0 °C to 40 °C,
humidity 35 uR% to 95 uR%
PC ABS UL94V0 self-extinguishing
3 PNP/NPN configurable as analogue for potentiometers.
(max 28 Vdc in PNP mode)
2 relays 5A resistive charge
30mA(at 24 VAV supply),40 mA(at 24 VDC supply),60
mA (at 110 to 230 VAC)
Rechargeable battery, approx. 7 days autonomy
24 to 230 VAC/VDC +/-15 % 50/60 Hz/2W
Front panel: IP65 (with gasket) ,
Box: IP30, Terminal blocks: IP20
.
Report the activation of Q1
MEANING
Report the activation of Q2
LED
123
123
Report serial transmission by the TA327401
123
MANUAL RATE METER
TA327401
MANUAL RATE METER
TA327401